Sign in →

Test Code LAB90 Glycated Hemoglobin (A1c)

Useful For

Monitoring long-term blood glucose control in individuals with diabetes mellitus

Methodology

Turbidimetric Inhibition Immunoassay (TINIA)

Performing Laboratory

Bozeman Health Laboratory Services

Specimen Requirements

Specimen Type: Whole blood

Container/Tube: Lavender top (EDTA)

Specimen Volume: 3 mL

Specimen Minimum Volume: 1 mL

Specimen Transport Temperature

Ambient/Refrigerated

Specimen Stability

Specimen Type Temperature Time
Whole blood (EDTA) Ambient 3 days
  Refrigerated 7 days
  Frozen 6 months

Reference Values

Normal: 4.0%-6.0%

 

Hemoglobin A1c Guidelines
5.7%-6.4% Pre-diabetic
≥ 6.5% Diabetic
< 7.0% Diabetic patient goal

 

Elevated levels of HbA1c suggest the need for more aggressive treatment of glycemia. The American Diabetes Association recommends that a primary goal of therapy should be a HbA1c of <7% and that physicians should reevaluate the treatment regimen in patients with HbA1c values consistently >8%.

Day(s) Test Set Up

Monday through Sunday

Test Classification and CPT Coding

83036